Inductively coupled plasma mass spectrometry (ICP-MS) separates ions by mass-to-charge ratio. Inductively coupled plasma optical emission spectrometry (ICP-OES) measures light at wavelengths characteristic of the elements. Both techniques can measure multiple elements. They differ in the concentration ranges they can cover and in the interferences that must be checked.

ICP-MS is often a candidate for trace and ultratrace measurements. ICP-OES is often practical for major constituents and higher concentrations. The choice depends on the required reporting limit, the concentration after preparation, the remaining sample matrix and the work needed to maintain reliable results.

Decision flow from the reporting requirement through dilution, matrix and interference checks
Enlarge figure
Use the sequence to shortlist a method, then verify it with representative samples.

Start with the reporting requirement in the original sample

  • Consider ICP-MS when the target must be quantified at very low concentrations after sample preparation and dilution.
  • Consider ICP-OES when the analyte is relatively concentrated and it is desirable to avoid excessive dilution.
  • When major and trace elements occur together, consider separate dilutions or separate methods.

The achievable limit of quantitation (LOQ) and working range depend on the element, matrix, blanks, preparation and instrument configuration. Verify them for the samples the method will receive.

Compare the two techniques

Conceptual ranges showing ICP-MS and ICP-OES with an overlap
Enlarge figure
The ranges are qualitative. There is no universal concentration cutoff between the techniques.

↔ 表が見切れている場合は、左右にスクロールしてご覧ください。↔ Scroll the table horizontally to see any hidden columns.

ComparisonICP-MSICP-OES
SignalIons formed in the plasma, separated by mass-to-charge ratioEmission from atoms or ions, separated by wavelength
Typical useTrace and ultratrace elementsMajor constituents and elements at higher trace concentrations
Sample considerationsMatrix load after dilution, contamination and blank controlHigher-concentration constituents, sample introduction stability and adequate sensitivity
Spectral interferenceIsobaric, polyatomic and doubly charged ionsOverlapping emission lines, background emission and stray light
Method settingsIsotope, internal standard, cell conditions and dilutionWavelength, background correction positions, viewing geometry and dilution
Routine checksSample introduction, interface, vacuum and cell performanceTorch, nebulizer, wavelength selection and background correction

Use the reporting limit to narrow the options, then check matrix load, interferences, maintenance and the likelihood of reruns. An instrument detection limit cannot stand in for a method LOQ that includes sample preparation. For the general principles, see US EPA Method 200.8 for ICP-MS, US EPA Method 200.7 for ICP emission spectrometry and Detection and Quantitation Limits.

1. Work back from the sample to the measured solution

  1. Identify the specification, action level or reporting limit in the original sample.
  2. Use sample mass and final preparation volume to calculate the equivalent concentration in the digest or extract.
  3. Account for any additional dilution needed to reduce the matrix load.
  4. Evaluate whether the candidate method can quantify that concentration with the required precision.

Even a sensitive instrument may struggle if sample preparation leaves the analyte too dilute. Reducing dilution can improve the concentration presented to the instrument, but it can also increase signal suppression, spectral interference and deposits in the sample introduction path. See How to Convert ICP Results from µg/L to mg/kg.

Worked example. Suppose the reporting limit for a solid is 0.10 mg/kg. You digest 0.50 g and make it up to 50 mL. The target concentration is 1.0 µg/L in the digest. After a further tenfold dilution to reduce salt, it is 0.10 µg/L. Compare this last value with the LOQ demonstrated in the sample matrix.

A 0.10 mg/kg target becomes 1.0 µg/L in the digest and 0.10 µg/L after a tenfold dilution
Enlarge figure
Illustrative calculation for a 0.50 g sample made up to 50 mL.

2. Identify the interferences that can affect the result

ICP-MS

An analyte signal can share the same nominal mass-to-charge ratio (m/z) with an isotope of another element, a polyatomic ion or a doubly charged ion. Polyatomic ions can arise from argon, oxygen, chloride or carbon in the plasma or sample. High matrix concentrations can also suppress or enhance the analyte signal. Check carryover after high-concentration samples.

Options include choosing another isotope, diluting or separating the matrix, using internal standards, applying a verified interference equation, or adjusting collision or reaction cell conditions. Which controls work depends on the analyte and matrix. US EPA Method 200.8, Section 4

ICP-OES

Emission from another element near the analyte wavelength can bias the result. Also check molecular bands, continuum background and stray light from strong lines. Inspect spectra from samples or interference check solutions to select the line and background positions. Options include an alternative wavelength, interelement correction, dilution or matrix separation. Verify that the selected conditions give suitable recovery. US EPA Method 200.7, Section 4

ICP-MS interference at the same m/z compared with overlapping lines in ICP-OES
Enlarge figure
In either technique, background or a second species can contribute to the analyte signal.

↔ 表が見切れている場合は、左右にスクロールしてご覧ください。↔ Scroll the table horizontally to see any hidden columns.

ObservationWhat to check in ICP-MSWhat to check in ICP-OES
High concentrations of matrix constituentsInternal standard response, suppression and interface depositsEmission intensity, background and sample introduction stability
A second signal near the analyteSame-m/z ions, isotope ratios and cell conditionsNearby lines, background positions and alternative wavelengths
A low-concentration sample follows a high oneCheck a rinse blank for carryoverCheck a rinse blank and inspect the introduction system

3. Assess the solution going into the instrument

Record total dissolved solids, major constituents, acid identity and concentration, and any organic solvent or residual carbon. Check for suspended particles, precipitates and undigested material. Consider whether the analyte could be lost by adsorption, volatilization or precipitation. Verify that dilution to a suitable sample load still leaves enough analyte to meet the reporting limit.

Trace ICP-MS analysis is particularly sensitive to contamination from reagents, vessels and the preparation environment. A heavy matrix can also produce deposits on the interface and affect ion transmission. ICP-OES is often considered for concentrated or high-matrix solutions, but the compatible levels of salts, acids and solvents depend on the torch, nebulizer, introduction system and method conditions.

4. Compare the whole workflow

Include blanks, calibration and quality control (QC), interference checks, rinse time, carryover checks and reruns. Consider maintenance of the torch, nebulizer and spray chamber. For ICP-MS, include the interface, vacuum system and collision or reaction cell conditions. Estimate the workload for changed dilutions, gas and standard consumption, consumables and waste, as well as training, troubleshooting and records.

When both methods meet the analytical requirements, compare routine sample throughput and the effort needed to maintain valid results.

Four examples from analytical practice

↔ 表が見切れている場合は、左右にスクロールしてご覧ください。↔ Scroll the table horizontally to see any hidden columns.

ScenarioInitial candidateFirst verification
Very low concentrations of hazardous elements in waterICP-MSReporting limit after preparation, blanks and mass interferences
A major constituent in a plating bathICP-OESCalibration range after dilution, background and introduction-system load
Trace elements in a saline solutionDepends on matrix and limit; consider dilution or separationConcentration after treatment and residual salt interference
Major and trace elements in the same digestSeparate aliquots and consider using both techniquesDilution history, contamination and recovery for each aliquot

Trace arsenic or lead in water

Convert the required reporting limit to the concentration in the measured solution. Consider ICP-MS if the requirement falls below the verified ICP-OES LOQ. For arsenic, check chloride-derived polyatomic ions along with the blank, internal standards and cell conditions. US EPA Method 200.8

Nickel as a major constituent of a plating bath

Dilute the bath into a suitable ICP-OES calibration range and verify the wavelength and background correction. Assess the load from acids, salts and surfactants before introducing the solution. Trace impurities in the same bath may need their own dilution and a different technique.

Trace elements in a saline solution

A hundredfold dilution lowers both salt and analyte concentrations by a factor of 100. Assess salt-related polyatomic ions, matrix suppression and deposits even when ICP-MS sensitivity is needed. If dilution leaves the analyte below the reporting requirement, evaluate matrix separation, preconcentration, a different sample introduction configuration or another analytical method. Check spike recovery and dilution linearity in the matrix.

Major and trace elements in one solid digest

Take separate aliquots of the digest. Use a larger dilution for major elements by ICP-OES and set the trace aliquot dilution according to the required ICP-MS sensitivity and matrix load. Record the dilution for each analyte and check contamination in the trace aliquot and calibration range in the major-element aliquot separately.

Checklist before selecting the method

  1. Confirm whether the applicable test method specifies a technique.
  2. Define the reporting limit and working range on the original sample basis.
  3. Calculate the expected concentration after preparation and dilution.
  4. Check which major constituents, acids, salts and organics remain.
  5. Confirm which interferences apply to the element and whether they can be managed.
  6. Decide whether trace and major elements need separate measurement conditions.
  7. Verify performance with blanks, spikes, or certified reference materials as appropriate.
  8. Include rinsing, maintenance, supplies and training in the operating plan.

If both techniques are permitted, compare LOQ, trueness, precision, dilution linearity, spike recovery, interferences and carryover using representative samples. Use the applicable method, instrument instructions and validation evidence to select the conditions.

Related guides