In high-performance liquid chromatography (HPLC), choose the stationary phase for retention and selectivity first. Then use column length, inner diameter and particle size to balance resolution, run time and pressure. If a phosphate-containing analyte gives a small peak area, compare the sample vial, column and instrument flow path before selecting low-adsorption hardware.

This guide starts with reversed-phase separations of small molecules and covers alternatives for polar compounds, charged species and biomolecules. Check the product instructions for pressure, temperature, pH, solvent and flow limits.

Three steps in HPLC column selection: stationary phase, dimensions and flow path
Enlarge figure
Choose selectivity first. Then select dimensions and wetted hardware.

Choose selectivity, dimensions and the flow path in that order

↔ 表が見切れている場合は、左右にスクロールしてご覧ください。↔ Scroll the table horizontally to see any hidden columns.

DecisionQuestion to askWhat to check
Separation mode and stationary phaseAre the analytes retained, and does the critical pair separate?Retention, elution order, resolution and solvent or pH compatibility
LengthWhat is the shortest column that meets the resolution requirement?Resolution, run time and pressure
Inner diameterWhat flow and injection volume suit the system and detector?Solvent use, injection volume and extra-column dispersion
Particle size and structureCan the system deliver the required efficiency within its pressure limit?Efficiency and the pressure ratings of the column and system
Protection and wetted hardwareWhere could particles accumulate or analytes adsorb?Filter and guard placement, recovery and peak shape

In a typical packed HPLC column, the stationary phase may be a bonded chemistry on silica particles or a porous material that itself provides the separation. In gas chromatography (GC), a capillary column can use a stationary-phase film on the capillary wall. Agilent HPLC column selection guide · Waters column hardware primer

Stationary phases and separation modes

Changing stationary phase or mobile phase can change elution order even if the dimensions stay the same. If a peak elutes close to the void time, try to increase retention. If a retained pair overlaps, investigate selectivity before increasing column length.

↔ 表が見切れている場合は、左右にスクロールしてご覧ください。↔ Scroll the table horizontally to see any hidden columns.

Mode or phaseTypical starting pointWhat to check
Reversed-phase C18 (octadecylsilane, ODS)Many small organic compoundsBase particle, bonding density, endcapping and pH range
Reversed-phase C8 or C4Excessive retention on C18, or screening for peptides and proteinsSelectivity and, for large molecules, pore size
Phenyl, phenyl-hexyl or pentafluorophenyl (PFP)Aromatic compounds or a pair that coelutes on C18Aromatic and dipolar interactions and the actual elution order
Polar-embedded or aqueous-compatible reversed phaseHighly aqueous eluents or selected basic analytesSpecified water content, wettability and peak shape
Normal-phase cyano (CN) or bare silicaDifferences in polarity that can be exploited in normal phaseSolvent compatibility, water content and the changeover procedure
Hydrophilic interaction liquid chromatography (HILIC)Polar species with weak reversed-phase retentionOrganic-rich eluent, equilibration and sample solvent. Partitioning into a water-rich layer and other interactions can contribute.
Cation or anion exchangeCharged molecules and proteinsAnalyte charge at the working pH, buffer concentration and sample ionic strength
Size-exclusion chromatography (SEC), including gel permeation chromatography (GPC) for polymersProtein aggregates or polymer molar-mass distributionsPore range, solvent and secondary interactions. Adsorption can distort a size-based separation.
Mixed-mode or chiral phasesDifficult pairs requiring more than one interaction, or enantiomersUse a mode and mobile phase suited to the specific separation

Examples by manufacturer and separation mode

Verify the exact phase, dimensions, particle specification, pore size and wetted materials. A single product family can contain several chemistries. Use the examples as starting points and compare performance with your own standards.

↔ 表が見切れている場合は、左右にスクロールしてご覧ください。↔ Scroll the table horizontally to see any hidden columns.

ModeExamplesWhat to verify
Reversed-phase C18ZORBAX Eclipse Plus C18 (Agilent), XBridge BEH C18 (Waters)Base particle, pH range and endcapping
Reversed-phase C18, Japanese suppliersL-column2 ODS (Chemicals Evaluation and Research Institute, Japan), Shim-pack GIST C18 (Shimadzu)Residual silanol activity and specifications
Reversed-phase C8 and C4Kinetex C8 (Phenomenex), Hypersil GOLD C4 (Thermo Fisher Scientific)Different bonded phases. Check pore size for proteins.
Reversed-phase C8, Japanese suppliersL-column2 C8 (Chemicals Evaluation and Research Institute, Japan), Shim-pack GIST C8 (Shimadzu)Particle and endcapping
Phenyl and PFPKinetex Phenyl-Hexyl, Kinetex F5 (Phenomenex)Each ligand gives a different selectivity
Phenyl and fluorophenyl, Japanese suppliersL-column2 C6-Phenyl (Chemicals Evaluation and Research Institute, Japan), Shim-pack GIST PFPP (Shimadzu)Phenyl-hexyl and pentafluorophenylpropyl ligands
Polar-embedded or aqueous-compatible reversed phaseAcclaim PolarAdvantage II (Thermo Fisher Scientific), Atlantis T3 (Waters)The polar-embedded phase and aqueous-compatible C18 use different designs
Aqueous-compatible reversed phase, Japanese supplierShim-pack GIST C18 AQ (Shimadzu)Permitted aqueous composition
Normal-phase cyano or silicaSpherisorb CN and Silica (Waters)Cyano-bonded or bare silica and the required solvent conditioning
HILICXBridge BEH Amide (Waters), SeQuant ZIC-HILIC (Merck)Amide and zwitterionic chemistries
HILIC, Japanese supplierShim-pack GIST Amide (Shimadzu)Equilibration and sample solvent
Ion exchangeProtein-Pak Hi Res IEX (Waters)Cation or anion exchange and strong or weak ligands
Ion exchange, Japanese supplierShim-pack WAX and WCX (Shimadzu)Weak anion or weak cation exchange for the analyte’s charge
SEC or GPCTSKgel SEC (Tosoh), AdvanceBio SEC (Agilent)Aqueous or organic operation and pore range
SEC, Japanese supplierShim-pack Bio Diol (Shimadzu)Pore size relative to the protein or peptide
Mixed-modeAcclaim Mixed-Mode WAX-1 (Thermo Fisher Scientific)Reversed-phase and weak anion-exchange contributions
ChiralCHIRALPAK IA (Daicel)Selector chemistry and compatible solvents

Pore size is a separate choice. Large peptides or proteins may enter small pores only to a limited extent. Choose a pore range that suits the molecular size and mode. Agilent HPLC column selection guide

Swapping the stationary phase changes separation and can change peak order
Enlarge figure
Weak retention and coelution of retained peaks call for different adjustments.

Column length, inner diameter and particle size

Length. With comparable packing and conditions, a longer column can provide more theoretical plates and higher resolution. In a simple comparison at constant plate height, selectivity and retention factor, resolution scales with the square root of length. Doubling length from 100 to 200 mm increases resolution by about 1.4 times, while increasing run time, pressure and solvent consumption at the same flow rate.

Inner diameter. At the same linear velocity, flow scales approximately with cross-sectional area. Moving from 4.6 to 2.1 mm gives (2.1 / 4.6)² ≈ 0.21. A flow of 1.0 mL/min therefore scales to about 0.21 mL/min as a starting point. Scale injection volume to suit column volume and minimise extra-column dispersion from tubing, fittings and the detector cell. With liquid chromatography–mass spectrometry (LC–MS), check flow and ionisation after the transfer.

Particle size. Smaller particles can improve efficiency or shorten the analysis. For otherwise comparable packed columns at the same linear velocity, the pressure drop can scale approximately with the inverse square of particle diameter. Changing from 5 to 2.5 µm gives roughly four times the pressure drop under that approximation. Packing structure, solvent viscosity, temperature and the rest of the flow path also matter. Compare superficially porous and fully porous particles using measured performance and pressure as well as nominal particle size. Waters primer on particle size and pressure

↔ 表が見切れている場合は、左右にスクロールしてご覧ください。↔ Scroll the table horizontally to see any hidden columns.

ChangePotential benefitCheck at the same time
100 to 150 mm lengthMore resolution for a close pairPressure, run time and gradient time
4.6 to 2.1 mm inner diameterLower flow and solvent consumptionInjection volume, tubing diameter, detector volume and response
5 to 3 µm particlesHigher efficiency or a faster runSystem and column pressure limits and filter loading
Resolution, flow and pressure trends when column dimensions change
Enlarge figure
The values illustrate simplified comparisons under otherwise comparable conditions.

When transferring a gradient method, also consider the gradient time, injection volume and dwell volume, the volume between the point of mobile-phase mixing and the column inlet. Recheck the critical pair, peak shape, quantitative result and system suitability. Waters gradient method scaling example

Guard columns, in-line filters and sample filters

↔ 表が見切れている場合は、左右にスクロールしてご覧ください。↔ Scroll the table horizontally to see any hidden columns.

ComponentMain functionWhat to check
Sample filterRemoves particles before injectionAnalyte adsorption, extractables and solvent compatibility. At trace levels, run a filter blank. Syringe filter selector (Japanese)
Mobile-phase filterReduces particulates entering the instrumentFiltration, degassing and conditions that prevent precipitation
In-line filterTraps particles immediately upstream of the column assemblyCheck it first if pressure rises. Dissolved, strongly retained contaminants can pass through.
Guard columnTraps particles and strongly retained contaminants before the analytical columnChoose suitable chemistry and diameter, and check the added pressure, volume and retention effects

Use appropriately prepared samples with a guard column. Extra guard length or a different phase can change selectivity and retention. Depressurise the system before checking for restrictions. Work from the filter to the guard and then the analytical column. Follow the column instructions for any reverse flushing. Thermo Fisher Scientific guide to guard columns

Sample filtration and the locations of the in-line filter, guard and analytical column
Enlarge figure
Filters catch particles. A guard can also retain dissolved contaminants.

A delay column can separate mobile-phase background from a sample peak

In the retention-gap method described for some liquid chromatography systems, a delay column, also called an isolator column, is installed between the pump or mixer and the injector. It delays contaminants from the mobile phase or upstream solvent-delivery system. The injected sample bypasses the delay column, so the same compound can appear at different times depending on its source.

A pre-injector delay column postpones the mobile-phase background relative to the injected sample
Enlarge figure
Only contaminants entering upstream of the injector pass through the delay column.

↔ 表が見切れている場合は、左右にスクロールしてご覧ください。↔ Scroll the table horizontally to see any hidden columns.

StreamPath and effect
Mobile phasePump and mixer, delay column, injector, analytical column, then detector
Upstream background contaminantTemporarily retained in the delay column before reaching the analytical column
Injected sampleEnters downstream of the delay column and reaches the analytical column directly

Tsuda and colleagues compared 50 and 100 mm delay columns for trace perfluorobutanoic acid (PFBA) and found 100 mm more effective for separating the background and sample peaks in their setup. Choose dimensions by checking the blank and target peaks in your own method. The added column changes pressure and can affect gradient delay. Check for interference with other analytes and continue to reduce contamination at its source. Contamination introduced at or after the injector bypasses the delay column. Tsuda et al., conference abstract (Japanese) · Waters example of isolator-column placement

If a strong sample solvent causes peak fronting, splitting or broadening, check the injection volume, sample solvent and initial mobile phase. A guard column is normally installed downstream of the injector to protect the analytical column, and a trap column can concentrate the injected sample. The pre-injector delay column addresses upstream background.

Low-adsorption columns and bioinert flow paths

A low-adsorption column can use treated wetted surfaces on the tube wall and the frits that retain the packing. A bioinert flow path also addresses surfaces elsewhere in the instrument, such as the injection assembly, tubing and valves. For either description, check the actual wetted components, materials and treatments. Waters comparison of metal-sensitive analyte recovery across LC systems · Agilent bio-inert system specifications

Phosphate-containing nucleotides and oligonucleotides, and some other metal-sensitive analytes, can give small areas, tailing or a low first-injection response after contact with metallic surfaces. Losses in the vial or strong retention on the stationary phase can also reduce the observed area. Waters nucleotide application study · Waters vial adsorption study

Conventional metal and low-adsorption flow paths compared
Enlarge figure
Check the full path from the sample vial to the column and detector.

How to investigate a low peak area

Use a standard near the reporting limit and change one component at a time. Compare areas, tailing and first-injection versus repeat-injection response. Run a blank after a high-concentration sample to check carryover.

↔ 表が見切れている場合は、左右にスクロールしてご覧ください。↔ Scroll the table horizontally to see any hidden columns.

ObservationComparison to runHow to interpret the result
Area falls as the sample waits in the vialSplit the same standard into conventional and low-adsorption vials. Compare fresh and held aliquots.Look for vial effects and check sample stability, degradation peaks and evaporation.
The first injection gives a smaller area than repeated injectionsCompare a conventional and a low-adsorption column with matching phase and dimensions. Use comparable equilibration for both.If the initial deficit is smaller on the low-adsorption column, its wetted surfaces are a plausible contributor.
Changing to a low-adsorption column has little effectKeep the column and vial fixed and check the injection system and tubing.Also consider strong stationary-phase retention and ion suppression in LC–MS.

Suppose the first injection gives 60% of the stable repeat-injection peak area on a conventional column and 95% on a comparable low-adsorption column. Each percentage is the first area relative to the stable area on that same column. A matrix spike recovery experiment answers a separate question. Before adopting the column, check repeatability near the reporting limit, peak shape and recovery in real samples. A change in separation or mobile-phase composition at elution can also change detector response. Waters first- and repeat-injection comparison for oligonucleotides · HPLC peak troubleshooting

Five practical column-selection examples

↔ 表が見切れている場合は、左右にスクロールしてご覧ください。↔ Scroll the table horizontally to see any hidden columns.

SituationWhat to try firstWhat to measure
Two impurities overlap on C18Adjust pH, organic solvent or gradient. Screen phenyl or PFP if more selectivity is needed.Resolution of the critical pair and reproducibility
A polar analyte has little reversed-phase retentionTry HILIC or ion exchange as appropriate. Review sample solvent and equilibration.Retention factor, peak shape and blank
An LC–MS method uses too much solventEvaluate 2.1 mm instead of 4.6 mm and optimise flow, injection volume and tubing.Response, resolution and pressure
Protein or nucleic acid recovery is lowCheck the pore range and mode. Test standards on comparable low-adsorption hardware.Recovery, tailing and first- versus repeat-injection area
Mobile-phase background overlaps a sample peak in a method for per- and polyfluoroalkyl substances (PFAS)Locate the contamination source and consider an upstream delay column.Separation of blank and sample peaks, recovery and added pressure

Before ordering, record the analytes, matrix, critical resolution, system pressure limit, detector and mobile phase. After installation, keep the part number, lot, dimensions, phase, pore size and first-use date. Record filter and guard replacements and save initial pressure, retention times and standard chromatograms. Choosing chromatographic vials and caps (Japanese)

How to use the references in your own method

↔ 表が見切れている場合は、左右にスクロールしてご覧ください。↔ Scroll the table horizontally to see any hidden columns.

SourceWhat it tells youWhat to check in your method
Product specification or manualPart number, dimensions, phase and operating limitsSeparation and recovery for your own sample
Manufacturer application notePerformance for a stated instrument, analyte and methodMethod transfer, matrix effects and suitability
Technical primerPrinciples of pressure, scaling and extra-column dispersionThe actual limit and performance of a given product

Even two C18 columns with the same dimensions can differ in substrate, pore size or surface treatment. Compare resolution of the critical pair, pressure and quantitative results before treating them as interchangeable. For detection choices, see the companion guide. HPLC detector selection